We use cookies to store information that makes it easier for you to use our website - for example, which language you want to use. These cookies do not store any personally identifiable information. You can set your browser to block or alert you about these cookies, but some parts of the website will not then work.

By continuing to use the website, you are agreeing to our Cookies Policy
Hilditch Auctions running_auctions_safely
Password: Soggetto a IVA Prezzo di riserva raggiunto

Maquet Servo-i Ventilator

Fisher Paykel Humidifier
SN: 091013062826
DOM: 10-2009

Air Hose
O2 Hose

Numero di serie:
44904

Ore di utilizzo:
5355.0

Apparecchiatura associata:
Fisher & Paykel Respiratory Humidifier Heater MR850WEK with heater wire cable/thermister

Accessori:
Air Hose
O2 Hose
2 x Battery packs
CO2 module with main stream sensor
Aeroneb Pro with mains adapter
Expiratory Cassette
User Manual

Software:
Version v8.0

Relazione sulle condizioni:
The Servo i power up and completed system initialisation. The control knobs, key pad and touch screen are working fine. Pre-use check was carried out and all tests passed. The system was tested with medical air and test lung, in volume control mode. Good ventilation cycles were delivered with the tidal volume (Vt) and breath per minute (BPM) corresponding to the measured minute volume (MV). O2 sensor, flow sensor and exp. cassette are in good working order. CO2 monitor is in good working order. Patient category: adult and infant. Vent modes: Pressure control, Volume control. PRVC, PS/CPAP, Volume Support, SIMV (Vol. Contr.) + Pressure Support, SIMV (Press. Contr.) + Pressure Support, SIMV (PRVC) + Pressure Control & NAVA. Installed options: Automode, Open Lung Tool, NIV, Software Version 3, Nebuliser, CO2 Analyser & NAVA. Module: CO2 & Edi Module.
The humidifier power up with function keys and display working, not fully tested. Ventilator is in good physical and cosmetic condition. Battery packs are working fine.
Total Operating Time: 5355 hours.
Ubicazione:
Malmesbury, Wiltshire, UK

Visualizzazione:
By appointment with the Auctioneer

Rimozione:
By Friday 6th August 2021

For every £100 you spend you will pay £20 Buyers Premium plus £4 VAT on that Premium [non-refundable] and on VAT registered lots £20 Hammer VAT [refundable upon proof of export].
All new and used medical, catering and general equipment lots are sold as seen. Where EST DOM’s (estimated dates of manufacture) are listed these are for guidance only. Any views expressed are for guidance only - please refer to our conditions of sale.

Printable Lot Details

Termina ven 30 lug 2021 | 12:32:55

Chiusura offerta POSTICIPATA

Numero lotto: 4  

Numero di vendita: 52218

Medicale | Asta online a tempo

Valuta vendita: £ UK Pound Sterling