We use cookies to store information that makes it easier for you to use our website - for example, which language you want to use. These cookies do not store any personally identifiable information. You can set your browser to block or alert you about these cookies, but some parts of the website will not then work.
By continuing to use the website, you are agreeing to our Cookies Policy
You are viewing our website in English
Hilditch Website - Select Language
Hilditch Website - Select Language
We’ve noticed you are in the UK. Do you want to view our website in English or change to English?
GE C1-5 Convex Array Transducer SN: 182409YP3 DOM: 08/2015
Sony Digital Graphic Printer UP-D898MD
Serial Number:
E72379
Manufacture Date:
11 / 2018
Software:
Version EC340 (19.0.3)
Condition Report:
The ultrasound system powers up and completes the initialisation process successfully. The LCD monitor, control panel, trackball and touch screen interface are all working accordingly. The motorised height adjustment and horizontal positioning adjustment of the control panel are in good working order and the electromagnetic brakes are also operational. Patient registration and image storing options are working accordingly and the printer is also in good working order. The functional system modes include 2D, Color, PD, PW, CW and M-mode. 3D and 4D modes are also functional and the HDlive option is also available. The transducer is working accordingly, tested with phantom and the system produced clear images with good resolution and greyscale and accurate caliper measurements. The transducer has some defective wires within the cable which produce interference in CW mode. The ultrasound unit has a few small crack lines on the edge of the control panel (near the handle). Otherwise the unit is in good physical condition with minor signs of wear.
Location:
Malmesbury, Wiltshire, UK
Viewing:
By appointment with the Auctioneer
Removal:
All items to be collected by Wednesday 29th March at 3pm. Items not collected by this date will incur late collection charges
For every £100 you spend you will pay £22.50 Buyers Premium plus £4.50 VAT on that Premium [non-refundable] and on VAT registered lots £20 Hammer VAT [refundable upon proof of export]. All new and used medical, catering and general equipment lots are sold as seen. Where EST DOM’s (estimated dates of manufacture) are listed these are for guidance only. Any views expressed are for guidance only - please refer to our conditions of sale.
In order to comply with new EU Privacy rules (GDPR), we need to ask you to read and accept our Terms and Conditions and Privacy Policy before continuing to use the website.